THAILAND · MEDICAL DEVICES

Thai medical device licence expiry and cessation stock reporting: how the 30, 60 and 15 days apply

The Thai FDA’s new announcement on reporting remaining medical device stock after business cessation or non-renewal has applied since 8 July 2026. For overseas manufacturers, the first step is to establish which Thai licence or certificate is ending and who holds the stock, and only then plan the reporting, sale or other disposal. “There are 60 more days to sell after expiry” does not apply to every product or every licence situation.

The announcement was signed on 15 December 2025, published in the Royal Gazette on 9 January 2026, took effect 180 days after publication, and replaces the 2021 (B.E. 2564) announcement on the same subject. The deadlines themselves come from Sections 36–38 of the Medical Device Act; the forms and submission method follow the new announcement. (Sources: announcement in the Royal Gazette; Thai FDA announcements page)

First identify which licence or certificate is ending

A Thai partner may hold a manufacturing or import establishment registration, product licences, detailed notifications or listing certificates, and sale licences for specific devices. Each document does something different and triggers different reporting. Ending a distribution agreement with the manufacturer does not in itself mean the authority has cancelled a certificate or accepted a transfer.

SituationCheck firstKey action
Voluntary cessation of businessThe actual date operations stopped, and the establishments and certificates involvedNotify cessation within 30 days under Section 36 (LP1) and return the certificates; licence and detailed-notification holders also report remaining quantities and storage locations.
Not renewed, or renewal refusedThe expiry date or the date of the refusal decisionReport remaining quantities and storage locations within 30 days under Section 37 (LP2).
Devices that require a sale licenceWhether the device is designated under Section 6(3) as requiring a Section 24 sale licenceAlso check the Section 38 disposal recipients and period, and the report due once it ends (LP3).
Only changing agent or distributorWhether the licence holder changes and whether certificates remain validArrange the registration and authorisation handover first; terminating a contract does not by itself trigger every cessation rule.

Where a suspension, revocation or other specific regulatory order is involved, check the operating restrictions in the actual decision first. The cessation and non-renewal procedures described here do not authorise continued supply of such products.

How the 30-day, 60-day and 15-day periods work

The deadlines are set by the Medical Device Act; the new announcement sets the matching forms and how to submit them. We suggest the Thai licence holder keeps a log of event dates and links each submission receipt to the stock records. The manufacturer supplies shipment and batch data, but this does not replace the filing by the local responsible party.

ProcedureLegal basisDeadline and start dateForm and key points
Cessation noticeSection 36Within 30 days of cessationLP1. Notify in writing and return the establishment registration, licence or notification certificate; licence and detailed-notification holders state remaining quantities and storage locations.
Non-renewal stock reportSection 37Within 30 days of expiry or of the refusal to renewLP2. Applies where an establishment registration, licence or detailed notification is not renewed or renewal is refused.
Disposal periodSection 38, paragraph 1Within 60 days of cessation, licence expiry or refusal to renewSection 24 sale licensees only; sale to another licensee or a person the authority considers appropriate. The authority may extend the period at its discretion; there is no automatic extension.
Report after the disposal periodSection 38, paragraph 2Within 15 days after that period endsLP3. Any such devices still held may no longer be sold; their quantity and storage location must be reported.

Under Clause 6 of the new announcement, filings are made electronically as the main route; where that is not yet possible, they are submitted at the Thai FDA or another place the Secretary-General designates in the Royal Gazette. Clause 7 provides that cessation notices filed before the announcement took effect are treated as made under it. Note also that Section 36, as amended, provides that if a business ceases without notifying, the establishment registration, licence, detailed notification or listing certificate is still deemed expired from the cessation date. Not notifying does not keep the certificates alive; it only leaves the stock reporting without a proper basis. (Source: Medical Device Act (No. 2) B.E. 2562, Section 25)

The 60 days are a specific disposal arrangement, not a substitute for restocking or renewal

Section 38 of the Medical Device Act applies to holders of a Section 24 sale licence: after cessation, licence expiry or refusal to renew, they may sell their remaining devices of that kind within 60 days to another licensee or to a person the authority considers appropriate. It does not give every expired product a general right to continue being sold to any customer. (Source: Medical Device Act, Section 38; original Gazette pp. 41–42)

A Section 24 sale licence is only required for devices the Ministry of Public Health has designated under Section 6(3) as needing a licensed seller. Paragraph 4 of Section 24 also treats manufacturers and importers holding a licence or notification under Section 17 or 19 as sale licensees for the devices they make or import. So whether Section 38 applies depends mainly on whether the device is in a category that requires a sale licence, not just on whether the local partner holds a separate sale licence certificate. For devices that do not require a sale licence, Section 38 does not provide this 60-day arrangement; how remaining stock is handled should be confirmed with the Thai FDA by the local licence holder for the specific case.

When preparing a disposal plan, list the products, quantities, intended recipients and their eligibility separately. Do not treat disposal of remaining stock as resuming import, continuing manufacture or replenishing stock, and do not assume that a commercially agreed “buyer” is automatically a recipient the law allows.

Before changing agent, check certificates and warehouse records together

Before deciding not to renew or to end a partnership, we suggest building a stock list that can be traced item by item, covering at least the device name, licence or notification number, lot/serial number, quantity and actual storage address; these match the columns of the LP3 form (device name, licence/notification number, lot or serial number, quantity, storage location). The LP3 form also has a “no remaining devices” box; whether a nil return is always required is worth confirming with the Thai FDA. Manufacturers can add expiry dates, orders and the intended disposal route for commercial planning.

The stock count should not stop at the outgoing agent’s own warehouse. Establish who physically holds the goods, whether there is third-party warehousing or channel stock, and which records the local licence holder can obtain. Ownership of the products, regulatory responsibility and physical possession may sit with different parties, so the handover agreement should allocate data provision, regulatory filings and custody separately.

We suggest writing the following into the exit plan: the last permitted shipment date, eligibility of disposal recipients, who files each report, handling of remaining and returned stock, and contacts for complaints and recalls. Whether a certificate transfer or a new application by another party can provide continuity needs a separate assessment; on changing the licence holder, see MediPass’s guide to Thai medical device licence transfer.

Frequently asked questions

Once the 30-day report is filed, can we keep selling for 60 days?
Filing alone does not get you there. The 30 days are the deadline for the cessation notice or stock report; the 60 days are the Section 38 disposal arrangement for devices that require a sale licence, and both the device category and the recipient must be checked. The two procedures have different conditions.

If stock remains after 60 days, can we report again and keep selling?
No. Section 38 expressly prohibits selling such devices once the period has ended and requires their quantity and storage location to be reported within 15 days. Filing LP3 does not grant a new sale licence; any further handling must be arranged as the authority requires.

Which devices require a Section 24 sale licence?
Those designated by the Ministry of Public Health by announcement under Section 6(3) of the Medical Device Act; not every medical device needs one. Check the current announcements product by product rather than inferring from the risk class.

If we change distributor but the licence holder stays the same, do we need to notify cessation?
First confirm the local operating and certificate status. Simply changing a commercial distribution partner is not the same as the licence holder ceasing business; if the holder’s establishment registration or product certificates are ending as well, check which of the procedures above applies.

Official references

  1. Thai Food and Drug Administration (Thai FDA), Announcement on criteria, methods and conditions for notifying business cessation and the quantity and storage location of remaining medical devices upon cessation, non-renewal or refused renewal… B.E. 2568, Clauses 1–7 and forms LP1–LP3; signed 15 December 2025, Royal Gazette Vol. 143, Special Part 7 Ngor, pp. 36–37, published 9 January 2026, effective 8 July 2026, medical.fda.moph.go.th (PDF)
  2. Thai FDA Medical Device Control Division, medical device announcements page (the above announcement and its implementation), medical.fda.moph.go.th
  3. Medical Device Act B.E. 2551 (2008), Section 6(3), Section 24 and Sections 36–38; Royal Gazette Vol. 125, Part 43 Kor, 5 March 2008 (Sections 36–38 at pp. 41–42 of the original; pp. 17–18 of the consolidated PDF linked here), cms-media.fda.moph.go.th (PDF)
  4. Medical Device Act (No. 2) B.E. 2562 (2019), Section 25 (replacing Section 36, adding listing certificates and the consequence of not notifying cessation); Royal Gazette Vol. 136, Part 56 Kor, 30 April 2019, pp. 196–197, catalog.fda.moph.go.th (PDF)

Prepared from official sources verifiable as of 10 July 2026 for manufacturers’ project planning; the handling of remaining stock in a specific case must be confirmed against the certificate type, the product and the authority’s decisions.

Leave a Comment

Your email address will not be published. Required fields are marked *

Chat on WhatsAppWhatsApp 咨询