VIETNAM · MEDICAL DEVICES
Vietnam Medical Device Post-Market Duties: Events, Recalls, Records
A European manufacturer holds a Vietnamese circulation number for a family of Class D implants. After a fracture incident involving the same batch in Germany, headquarters instructs every country holder to “report to the authority within 48 hours”. The Vietnamese holder opens the decree and finds no 48-hour figure, but a different logic: suspend the batch, notify the Ministry of Health (MOH) in writing, and complete an investigation within 30 days of the warning.
Vietnam’s post-market obligations are not organised as a reporting-deadline table but as the holder’s continuing duties under Article 74 and the event-handling procedures of Articles 33 to 39a of Decree 98/2021/ND-CP: there is no fixed adverse-event reporting deadline, no annual report and no renewal, but each step carries a statutory action and an MOH decision timeline.
If the circulation number never expires, what does the holder have to maintain?
Article 21(3) provides that a circulation number for medical devices and IVDs is valid indefinitely; only emergency-issued numbers carry a validity period set by the Minister of Health. Vietnam therefore has no renewal or annual retention procedure. (Source: Integrated document 08/VBHN-BYT, 2026-03-06, §21(3))
Indefinite validity does not mean no maintenance duties. Among the holder obligations in Article 74(3), those most relevant to post-market work are: (c) establish or contract a warranty facility; (d) keep tracking records and ensure traceability; (e) issue prompt warnings about risks; (g) promptly stop circulation, notify the parties concerned and repair or recall defective devices, bearing any destruction cost; (k) keep the free-sale certificate, authorisation letter and warranty eligibility document valid throughout circulation; (n) provide purchasing healthcare facilities with the management file under Article 33(4); and (o) declare prices. (Source: Nghị định 98/2021/NĐ-CP, 2021-11-08, Điều 74 khoản 3; Integrated document 08/VBHN-BYT, 2026-03-06, §74(3))
Point (k) is the one most often overlooked: if the CFS or authorisation letter lapses, the number does not fall away, but the holder is in breach and will be asked for valid documents at the next change filing, tender or inspection.
Which post-market records does Article 33 require?
Article 33 requires the holder to organise traceability and keep at least: the registration documents in paper form (authorisation letter, warranty eligibility confirmation, CFS or marketing authorisation); distribution records (a representative office may have its authorised importer keep them); and quality management documents, including the certificate of origin, batch quality certificates, inspection records for devices listed in Article 55(1), repair and maintenance documents, the Vietnamese IFU and warranty facility details. Because Articles 34 and 35 operate batch by batch, distribution records determine whether a recall can be executed at all. (Source: Integrated document 08/VBHN-BYT, 2026-03-06, §33)
After a safety warning or an incident, what must the holder do and by when?
Article 34 covers warnings about devices that pose a serious threat to public health or may cause death. On such a warning from a Vietnamese or international authority, the holder must notify user facilities, complete an investigation within 30 days (or explain to the MOH why not, with the safety measures taken), suspend the batch, notify the MOH, distributors and users in writing, and prepare a repair or recall plan, reporting on completion. For a repairable defect the MOH issues a suspension decision within 3 working days of the notice; after the holder repairs and reports, the MOH decides within 20 working days whether to lift the suspension. For an irreparable defect the MOH orders recall of the whole batch, enforced under administrative-violation law if the deadline is missed. (Source: Integrated document 08/VBHN-BYT, 2026-03-06, §34(1)–(4))
Article 35 covers adverse events that have already harmed users. Where the event caused a serious threat or death, the holder must post a notice on its website (if any), notify the MOH, traders and users in writing, suspend the batch, investigate and report; where it caused serious harm without death, the holder notifies the MOH and reports after investigation. Defective devices are then handled under Article 34(3)–(4). The decree sets no hours or days for the first notification; “promptly” is the only standard. (Source: Integrated document 08/VBHN-BYT, 2026-03-06, §35)
| Situation | Holder action | MOH timeline | Basis |
|---|---|---|---|
| Authority warning (serious threat / risk of death) | Notify facilities; investigate within 30 days; suspend batch; written notice to MOH | Suspension decision 3 working days; lifting decision 20 working days | §34 |
| Event causing serious threat or death | Website notice; written notice to MOH, traders, users; suspend batch; investigate and report | Per §34(3)–(4) | §35(1) |
| Event causing serious harm, no death | Notify MOH; investigate, report, repair or recall | Per §34(3)–(4) | §35(2) |
| Three batches recalled under one number | — (voluntary recalls excluded) | Revocation considered within 5 working days and published | §38(2), §39 |
What forms can a recall take, and when does it escalate to revocation?
Article 36 lists the remedies for defective devices: instructions to fix the defect, repair, replacement with an equivalent device, or recall for re-export or destruction; recalls are either voluntary or mandatory. Article 38(2) provides that the circulation number is revoked where three batches of the device are recalled during its validity, except where the holder recalled voluntarily; other grounds include forged documents, quality inconsistent with the registration and, under clause 14 added by Decree 07/2023, a competent authority’s finding that a dossier component is unlawful. Under Article 39 the issuing authority considers revocation within 5 working days of receiving the inspection record, publishes the decision and notifies the holder, provincial health departments and customs; a holder may also apply for voluntary revocation. (Source: Integrated document 08/VBHN-BYT, 2026-03-06, §36, §38(2), §39; Nghị định 07/2023/NĐ-CP, 2023-03-03, Điều 1 khoản 6–8)
Article 39a, added by Decree 07/2023, settles what happens after revocation: devices already sold to healthcare facilities or users may continue in use until disposed of under law or expiry, except irreparably defective devices harmful to users under Article 34(4); unsold devices must be withdrawn from circulation and recalled by the holder. Revocation and recall are therefore separately assessed events. (Source: Nghị định 07/2023/NĐ-CP, 2023-03-03, Điều 1 khoản 8 (Điều 39a))
If the manufacturer stops production or is dissolved, how long can the product stay on the market?
Under Article 37, where the product owner ceases manufacturing, goes bankrupt or is dissolved, the device may remain on the market for at most 24 months from the declaration date, and the Vietnamese holder must commit to warranty or spare-part support for 8 years; where the holder itself is bankrupt or dissolved, the same limit applies and the distributor gives the commitment. The commitment and device list are filed through the portal within 60 days and the MOH replies within 15 working days; if continued circulation is refused, devices still on the market must be recalled, except those already sold to facilities or users. (Source: Integrated document 08/VBHN-BYT, 2026-03-06, §37; Nghị định 07/2023/NĐ-CP, 2023-03-03, Điều 1 khoản 4)
Are there periodic reports, and what does the portal publish?
The decree imposes no annual or periodic post-market report for ordinary devices. The only periodic report is in Article 74(4): entities trading, importing, exporting or transferring devices, raw materials or external quality-control materials containing narcotic substances or precursors must file an annual report with the MOH and the Ministry of Public Security before 15 January of the following year. (Source: Nghị định 98/2021/NĐ-CP, 2021-11-08, Điều 74 khoản 4)
The portal’s role is publicity: within 1 working day of issuance the MOH publishes the device, product owner, holder, warranty facility and intended purpose, and revocation decisions are published and relayed by provincial health departments. Hospitals check supplier credentials against it, so the portal record is the baseline for external consistency. (Source: Integrated document 08/VBHN-BYT, 2026-03-06, §31(6), §39(3)–(4))
On responsibilities, the 30-day investigation deadline sits with the holder while root-cause analysis is usually done by the manufacturer, so the holder agreement should fix the manufacturer’s response time. As a hypothetical example: for a global FSCA notice issued on 1 March 2027, the Vietnamese holder suspends the batch and notifies the MOH the same day, and the manufacturer must deliver its findings by 20 March for the report to be completed by 31 March; otherwise the holder must explain the delay and the safety measures taken.
Frequently asked questions
Does Vietnam have a 48-hour or 10-day adverse-event reporting deadline?
No. Articles 34 and 35 require only “prompt” notification and a 30-day investigation; the 3- and 20-working-day periods are the MOH’s own decision timelines.
Will a voluntary recall lead to revocation of the circulation number?
No. Article 38(2) expressly excludes voluntary recalls from the three-batch rule, but the recall must still be notified to the MOH and reported under Articles 34–36.
Must the holder file an annual post-market report with the MOH?
Not for ordinary devices. The only periodic report is the Article 74(4) annual report for devices and materials containing narcotic substances or precursors, due before 15 January.
What happens to devices hospitals have already bought when a number is revoked?
Under Article 39a they may continue in use until disposal or expiry, except irreparably defective devices harmful to users; unsold devices must be withdrawn and recalled.
Official references
- Ministry of Health, Integrated document 08/VBHN-BYT – Decree on medical device management (English translation, thuvienphapluat), 2026-03-06. https://thuvienphapluat.vn/van-ban/EN/The-thao-Y-te/Integrated-document-08-VBHN-BYT-2026-Decree-medical-device-management/700470/tieng-anh.aspx
- Bộ Y tế, Văn bản hợp nhất 08/VBHN-BYT (Vietnamese original, Government portal), 2026-03-06. https://datafiles.chinhphu.vn/cpp/files/vbpq/2026/3/08-vbhn-byt.pdf
- Bộ Y tế, Văn bản hợp nhất 05/VBHN-BYT – Nghị định quy định về quản lý thiết bị y tế, 2025-03-26. https://thuvienphapluat.vn/van-ban/The-thao-Y-te/Van-ban-hop-nhat-05-VBHN-BYT-2025-Nghi-dinh-quan-ly-thiet-bi-y-te-649819.aspx
- Chính phủ, Nghị định 98/2021/NĐ-CP về quản lý trang thiết bị y tế, Điều 74, 2021-11-08. https://hethongphapluat.com/nghi-dinh-98-2021-nd-cp-ve-quan-ly-trang-thiet-bi-y-te/dieu-74
- Chính phủ, Nghị định 07/2023/NĐ-CP sửa đổi Nghị định 98/2021/NĐ-CP, 2023-03-03. https://thuvienphapluat.vn/van-ban/The-thao-Y-te/Nghi-dinh-07-2023-ND-CP-sua-doi-Nghi-dinh-98-2021-ND-CP-quan-ly-trang-thiet-bi-y-te-557403.aspx
- Ministry of Finance, Circular 59/2023/TT-BTC prescribing fees in the medical sector, 2023-08-30. https://thuvienphapluat.vn/van-ban/thue-phi-le-phi/circular-59-2023-tt-btc-prescribing-fees-in-medical-sector-and-collection-transfer-management-579034.aspx
This article supports general compliance planning; the applicable route, documents and timelines must be confirmed against the product, its registration status and the authority’s case-specific requirements.